In this study, BrachyDose, a recently developed EGSnrc Monte Carlo code for rapid brachytherapy dose calculations, has been benchmarked by reproducing previously published dosimetry parameters for three brachytherapy seeds with varied internal structure and encapsulation. Calculations are performed for two 125 seeds (Source Tech Medical Model STM1251 and Imagyn isoSTAR model 12501) and one Pd103 source (Theragenics Model 200). Voxel size effects were investigated with dose distribution calculations for three voxel sizes: 0.1×0.1×0.1 mm3, 0.5×0.5×0.5 mm3, and 1×1×1 mm3. In order to minimize the impact of voxel size effects, tabulated dosimetry data for this study consist of a combination of the three calculations: 0.1×0.1×0.1 mm3 voxels for distances in the range of 0<r≤1 cm, 0.5×0.5×0.5 mm3 voxels for 1<r≤5 cm and 1×1×1 mm3 voxels for 5<r≤10 cm. Dosimetry parameters from this study are compared with values calculated by other authors using Williamson's PTRAN code and to measured values. Overall, calculations made with Brachydose show good agreement with calculations made with PTRAN although there are some exceptions.

Additional Metadata
Persistent URL dx.doi.org/10.1118/1.2400843
Journal Medical Physics
Citation
Taylor, R.E.P., Yegin, G., & Rogers, D.W.O. (2007). Benchmarking BrachyDose: Voxel based EGSnrc Monte Carlo calculations of TG-43 dosimetry parameters. Medical Physics, 34(2), 445–457. doi:10.1118/1.2400843